A novel technique to treat osteoporotic proximal humeral fractures with diaphyseal extension

Authors

  • Ahmed Fouad Abotaleb Faculty of Medicine, Alexandria University, Egypt

DOI:

https://doi.org/10.18203/issn.2455-4510.IntJResOrthop20252448

Keywords:

Osteoporotic fractures, Proximal humeral fractures, Comminuted fractures, Diaphyseal extension, Double plating, Double plating proximal humerus, Dual plating

Abstract

Background: Recent reports of failure of proximal humeral locked plate (PHILOS) in treating comminnuted proximal humeral fractures (PHF) with diaphyseal extension (DE) prompted attempts to either augment or add more fixation which is challenged with the proximity to important neurovascular structures.

Methods: Between February 2020 and May 2024, 29 patients with comminuted fractures of PHF-DE were treated at El-Hadra University Hospital with the addition of either a small locked dynamic compression plate (sLDCP) or a small locked reconstruction plate (sLRP) to the standard PHILOS plate, using the trans-deltoid approach.

Results: The average follow-up (FU) period was about 19.5 months. The average age of the patients was approximately 62.6 years. Most of the patients (72.4%) were females. All patients healed their bones without needing additional procedures like bone grafts. The average healing time was just over six months. The mean visual analogue scoring system (VAS) was 8.2. At the end of the FU, there was no significant change in the neck shaft angle (p=0.7). At the end of the FU, the average Constant-Murley shoulder score was noted as 86.6, the mean Simple Shoulder Score was 79.3%, and the average score for the shoulder rating at the University of California Los Angeles (UCLA) was 31.1. No iatrogenic neurovascular fractures occurred during the study.

Conclusions: Performing double plating on PHF-DE is safe. It does not increase the risk of shoulder impingement, harm to the blood supply of the humeral head, or nerve injuries caused by treatment.

Metrics

Metrics Loading ...

References

Court-Brown CM, Duckworth AD, Clement ND, McQueen MM. Fractures in older adults. A view of the future? Injury. 2018;49(12):2161-6. DOI: https://doi.org/10.1016/j.injury.2018.11.009

Katthagen JC, Grabowski S, Huber M, Jensen G, Voigt C, Lill H. Epidemiology and treatment reality of proximal humeral fractures at a level-1 trauma center. Obere Extremität. 2016;2(11):112-8. DOI: https://doi.org/10.1007/s11678-016-0362-4

Handoll H, Brealey S, Rangan A, Keding A, Corbacho B, Jefferson L, et al. The ProFHER (PROximal Fracture of the Humerus: Evaluation by Randomisation) trial-a pragmatic multicentre randomised controlled trial evaluating the clinical effectiveness and cost-effectiveness of surgical compared with non-surgical treatment for proximal fracture of the humerus in adults. Health Technology Assessment (Winchester, England). 2015;19(24):1. DOI: https://doi.org/10.3310/hta19240

Brorson S, Viberg B, Gundtoft P, Jalal B, Søren O-N. Epidemiology and trends in management of acute proximal humeral fractures in adults: an observational study of 137,436 cases from the Danish National Patient Register, 1996-2018. Acta Orthopaedica. 2022;93:750. DOI: https://doi.org/10.2340/17453674.2022.4578

Lander ST, Mahmood B, Maceroli MA, Byrd J, Elfar JC, Ketz JP, et al. Mortality rates of humerus fractures in the elderly: does surgical treatment matter? J Orthopaedic Trauma. 2019;33(7):361-5. DOI: https://doi.org/10.1097/BOT.0000000000001449

Patel AH, Wilder JH, Ofa SA, Lee OC, Savoie III FH, O’Brien MJ, et al. Trending a decade of proximal humerus fracture management in older adults. JSES Int. 2022;6(1):137-43. DOI: https://doi.org/10.1016/j.jseint.2021.08.006

Rischen R, Köppe J, Stolberg-Stolberg J, Freistühler M, Faldum A, Raschke MJ, et al. Treatment reality of proximal humeral fractures in the elderly-trending variants of locking plate fixation in Germany. J Clin Med. 2023;12(4):1440. DOI: https://doi.org/10.3390/jcm12041440

Thanasas C, Kontakis G, Angoules A, Limb D, Giannoudis P. Treatment of proximal humerus fractures with locking plates: a systematic review. J Shoulder Elbow Surg. 2009;18(6):837-44. DOI: https://doi.org/10.1016/j.jse.2009.06.004

Schliemann B, Wähnert D, Theisen C, Herbort M, Kösters C, Raschke MJ, et al. How to enhance the stability of locking plate fixation of proximal humerus fractures? An overview of current biomechanical and clinical data. Injury. 2015;46(7):1207-14. DOI: https://doi.org/10.1016/j.injury.2015.04.020

Gardner MJ, Weil Y, Barker JU, Kelly BT, Helfet DL, Lorich DG. The importance of medial support in locked plating of proximal humerus fractures. J Orthopaed Trauma. 2007;21(3):185-91. DOI: https://doi.org/10.1097/BOT.0b013e3180333094

Sun Q, Wu X, Wang L, Cai M. The plate fixation strategy of complex proximal humeral fractures. Int Orthopaed. 2020;44:1785-95. DOI: https://doi.org/10.1007/s00264-020-04544-7

Osterhoff G, Ossendorf C, Wanner GA, Simmen H-P, Werner CM. The calcar screw in angular stable plate fixation of proximal humeral fractures-a case study. J Orthopaed Surg Res. 2011;6:1-6. DOI: https://doi.org/10.1186/1749-799X-6-50

Newman J, Kahn M, Gruson KI. Reducing postoperative fracture displacement after locked plating of proximal humerus fractures: current concepts. Am J Orthop. 2015;44(7):312-20.

Stone MA, Namdari S. Surgical considerations in the treatment of osteoporotic proximal humerus fractures. Orthopedic Clin. 2019;50(2):223-31. DOI: https://doi.org/10.1016/j.ocl.2018.10.005

Choi S, Seo K-B, Kwon YS, Kang H, Cho C, Rho JY. Dual plate for comminuted proximal humerus fractures. Acta Orthop Belg. 2019;85(4):429-36.

Oppebøen S, Wikerøy AK, Fuglesang HF, Dolatowski FC, Randsborg P-H. Calcar screws and adequate reduction reduced the risk of fixation failure in proximal humeral fractures treated with a locking plate: 190 patients followed for a mean of 3 years. J Orthop Surg Res. 2018;13:1-8. DOI: https://doi.org/10.1186/s13018-018-0906-y

He Y, He J, Wang F, Zhou D, Wang Y, Wang B, et al. Application of additional medial plate in treatment of proximal humeral fractures with unstable medial column: a finite element study and clinical practice. Medicine. 2015;94(41):e1775. DOI: https://doi.org/10.1097/MD.0000000000001775

He Y, Zhang Y, Wang Y, Zhou D, Wang F. Biomechanical evaluation of a novel dualplate fixation method for proximal humeral fractures without medial support. Journal of orthopaedic surgery and research. 2017;12:1-10. DOI: https://doi.org/10.1186/s13018-017-0573-4

Choi S, Kang H, Bang H. Technical tips: dualplate fixation technique for comminuted proximal humerus fractures. Injury. 2014;45(8):1280-2. DOI: https://doi.org/10.1016/j.injury.2014.04.029

Theopold J, Marquaß B, Fakler J, Steinke H, Josten C, Hepp P. The bicipital groove as a landmark for reconstruction of complex proximal humeral fractures with hybrid double plate osteosynthesis. BMC Surg. 2016;16:1-6. DOI: https://doi.org/10.1186/s12893-016-0125-6

Theopold J, Schleifenbaum S, Müller M, Werner M, Hammer N, Josten C, et al. Biomechanical evaluation of hybrid double plate osteosynthesis using a locking plate and an inverted third tubular plate for the treatment of proximal humeral fractures. PloS One. 2018;13(10):e0206349. DOI: https://doi.org/10.1371/journal.pone.0206349

Wanner GA, Wanner-Schmid E, Romero J, Hersche O, von Smekal A, Trentz O, et al. Internal fixation of displaced proximal humeral fractures with two one-third tubular plates. J Trauma Acute Care Surg. 2003;54(3):536-44. DOI: https://doi.org/10.1097/01.TA.0000052365.96538.42

Hessmann MH, Korner J, Hofmann A, Sternstein W, Rommens PM. Angle-fixed plate fixation or double-plate osteosynthesis in fractures of the proximal humerus: a biomechanical study. Biomed Tech Biomed Engineering. 2008;53(3):130-7. DOI: https://doi.org/10.1515/BMT.2008.018

Michel PA, Raschke MJ, Katthagen JC, Schliemann B, Reißberg I, Riesenbeck O. Double plating for complex proximal humeral fractures: clinical and radiological outcomes. J Clin Med. 2023;12(2):696. DOI: https://doi.org/10.3390/jcm12020696

Doornink J, Fitzpatrick DC, Boldhaus S, Madey SM, Bottlang M. Effects of hybrid plating with locked and nonlocked screws on the strength of locked plating constructs in the osteoporotic diaphysis. J Trauma Acute Care Surg. 2010;69(2):411-7. DOI: https://doi.org/10.1097/TA.0b013e3181ec9417

Hackl M, Wegmann K, Taibah S, Burkhart KJ, Scaal M, Müller LP. Peri-implant failure in dual plating of the distal humerus-a biomechanical analysis with regard to screw and plate positioning. Injury. 2015;46(11):2142-5. DOI: https://doi.org/10.1016/j.injury.2015.08.015

Downloads

Published

2025-07-30

How to Cite

Abotaleb, A. F. (2025). A novel technique to treat osteoporotic proximal humeral fractures with diaphyseal extension. International Journal of Research in Orthopaedics, 11(5), 1019–1027. https://doi.org/10.18203/issn.2455-4510.IntJResOrthop20252448

Issue

Section