The accuracy of free-hand cup positioning: a CT based measurement in a South Indian hospital

Vetrivel Chezian Sengodan, S. Marimuthu, Raghupathy Ramasamy

Abstract


Background: In total hip arthroplasty, the restoration of normal hip center in acetabular reconstruction encourages the normal bio mechanics. The impingement, dislocation, and implant wear have increased the interest in accurate component placement in arthroplasty. This study is to determine the accuracy of the free hand technique in the acetabular cup placement in relation to native acetabulam using CT scan images.

Methods: This study was done in the Coimbatore Medical College and Hospital from June 2015 to March 2016. Twenty patients who have undergone total hip replacements were subjected to CT scan of pelvis preoperatively and postoperatively. The acetabular cup version and the inclination were measured and analyzed statistically.

Results: The mean acetabular version was 17.5◦±6.5◦ and the inclination was 46.6◦±9.3◦ pre operatively. Post operatively the mean acetabular version was 26.3◦±17.3◦ and the inclination was 41.7◦±10.9◦. With reference to the component version it ranged from 12.5˚ of retroversion to 62.5˚ anteversion with a mean of 26.3◦±17.3◦. The component inclination ranged from 15.6◦ to 58.4◦ with a mean of 41.7◦±10.9◦. On combining component version and inclination, only (50%) was within the safe zone.

Conclusions: Based on the CT finding acetabular version and inclination differs from individual to individual. During acetabular cup placement in total hip arthroplasty using free hand technique, our study shows that that inclination is better achieved and the anteversion is over targeted in most of the cases.


Keywords


Acetabular cup, Computed Tomography, Cup positioning, THA

Full Text:

PDF

References


Bargar WL, Jamali AA, Nejad AH. Femoral ante version in THA and its lack of correlation with native acetabular anteversion. Clin Orthop Relat Res. 2010;468:527-32.

Saikia KC, Bhuyan SK, Rongphar R, Anthropometric study of hip joint in northeastern population with computed tomography scan. Indian J Orthop. 2008;42:260-6.

Murray DW. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75(2):228–32.

Ghelman B, Kepler CK, Lyman S, Della Valle AG. CT outperforms radiography for determination of acetabular cup version after THA. Clin Orthop Relat Res. 2009;467:2362–70.

Reikeras O, Gunderson RB. Components anteversion in primary cementless THA using straight stem and hemispherical cup: a prospective study in 91 hips using CT-scan measurements. Orthop Traumatol Surg Res. 2011;97:615–21.

Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217-20.

Coventry MB, Beckenbaugh RD, Nolan DR, Ilstrup DM. 2012 Total hiparthroplasties: a study of postoperative course and early complications. J Bone Joint Surg Am. 1974;56:273-84.

Sanchez-Sotelo J, Berry DJ. Epidemiology of instability after total hip replacement. Orthop Clin North Am. 2001;32:543-52.

Bargar WL, Jamali AA, Nejad AH. Femoral anteversion in THA and its lack of correlation with native acetabularanteversion. Clin Orthop Relat Res. 2010;468:527-32.

Jolles BM, Genoud P, HoVmeyer P. Computer-assisted cup placement techniques in total hip arthroplastyimprove accuracy of placement. Clin Orthop Relat Res 2004;426:174–9.

Hirakawa K, Mitsugi N, Koshino T, Saito T, Hirasawa Y, Kubo T. Effect of acetabular cup position and orientation in cemented total hip arthroplasty. Clin Orthop Relat Res. 2001;388:135-42.

Leenders T, Vandevelde D, Mahieu G, Nuyts R. Reduction in variability of acetabular cup abduction using computer assisted surgery: a prospective and randomized study. Comput Aided Surg. 2002;7:99–106.

Bosker BH, Verheyen CC, Horstmann WG, Tulp NJ. Poor accuracy of freehand cup positioning during total hip arthroplasty. Arch Orthop Trauma Surg. 2007;127:375-9.

Saxler G, Marx A, Vandevelde D, Langlotz U, Tannast M, Wiese M, et al. The accuracy of free-hand cup positioning-a CT based measurement of vcup placement in 105 total hip arthroplasties. Int Orthop. 2004;28:198-201.

Digioia 3rd AM, Jaramaz B, Plakseychuk AY, Moody JE Jr, Nikou C, Labarca RS, et al. Comparison of a mechanical acetabular alignment guide with computer placement of the socket. J Arthroplasty. 2002;17:359-64.

Kalteis T, Handel M, Bäthis H, Perlick L, Tingart M, Grifka J. Imageless navigation for insertion of the acetabular component in total hip arthroplasty: is it as accurate as CT-based navigation? J Bone Joint Surg Br. 2006;88(2):163–7.